Overview

Ultrasonic, noncontacting sensor. Ideal for weirs and flumes in heavily contaminated streams
The 4110 Ultrasonic Flow Logger provides non-contact measurement of the flow over a weir or in a flume. The ultrasonic sensor is mounted above the flow stream, and transmits a sound pulse that is reflected by the liquid surface. Elapsed time between sending a pulse and receiving an echo determines the level in the channel. The 4110 calculates the flow from this level reading and its built in conversion for the specified primary device. Automatic compensation for air temperature ensures measurement accuracy.

Because its sensor does not contact the liquid, the 4110 provides long-term dependability with no scheduled maintenance. The Isco 4110 is not affected by chemicals or high concentrations of grease, suspended solids or silt in the flow.

All setup of the flow logger is performed using Flowlink Software. You can retrieve stored data on-site with a laptop PC or an Isco 581 Rapid Transfer Device. Isco Flowlink Software produces a variety of informative graphs and reports from stored data.
Features

• Large, versatile memory - stores over 100000 readings of flow, rainfall and sample data.
• Easy setup using Flowlink software (required)
• Easy to upgrade - non-volatile "flash" memory makes it easy to upgrade firmware without opening the enclosure.
• Rugged, submersible enclosure (NEMA 4X, 6 and IP67)
• Up to one year battery life using standard alkaline lantern batteries.
• Variable Blanking Distance eliminates false echo problems caused by obstructions such as manhole rungs or the top of a flume.

Benefits

Rugged, Dependable Sensor
The 4110 uses a single-head sensor sealed in a rugged, corrosion resistant enclosure. The sensor transmits a sound pulse which is reflected from the surface of the flow stream. The elapsed time between sending a pulse and receiving an echo determines the level in the channel.

Accurate Under Tough Conditions
To ensure maximum accuracy, a built-in sensor automatically compensates for changes in air temperature. To improve performance in the presence of turbulence and foam, the 4110 automatically adjusts amplifier gain in response to echo strength. And, Isco's Variable Blanking Distance feature eliminates false echo problems caused by obstructions such as manhole rungs or the top of a flume.

Applications

• Flow measurement in streams containing harsh chemicals, grease or suspended solids
• Sampler Pacing and Activation - a 4100 Series Flow Logger signals a sampler to collect flow proportioned samples, and can also trigger a sampler based on flow, rainfall, or time.
• Treatment capacity analysis
• Inflow and infiltration studies
• Combined sewer overflow studies
• Long-term river and stream gauging

Specifications

Part Number: 68-4110-001
Baud Rate 300, 1200, 2400, 4800 or 9600
Blanking Distance 0.6m to 3.7m (2 ft to 12 ft)
Cable 7.6m × 0.8cm (25 ft × 0.3")
Compensated Temperature -30°C to 60°C (-22°F to 140°F)
Data Partitions Maximum of 6
Data Points 50 level-flow rate points
Data Types Level, flow rate, rainfall or sample data
Dimensions (Length × Diameter) 17.5cm × 9.1cm (6.9" × 3.6")
Enclosure NEMA 4X, 6; IP67
Enclosure NEMA 4X, 6 (IP67)
Equations Two-term polynomial of the form K^1H^N1 ±K^2H^N2
Flumes Parshall, Palmer-Bowlus, Leopold-Lagco, trapezoidal, H
For Head Change <0.31m (<1.0 ft) Maximum error is ±0.006m (±0.02 ft)
For Head Change of 0.31m to 3.05m (1.0 ft to 10 ft) Maximum error is ±0.009m (±0.03 ft)
Frequency 40 kHz
Manning Formula Round, U-channel, rectangular, trapezoidal
Material Body and battery end are structural foam molded polystyrene. Connector end is Ryton® (polyphenylene sulfide).
Materials Sensor housing is Xenoy 6120; Cable jacket is Polyvinyl chloride (PVC)
Memory Capacity 230000 bytes; equal to over 700 days of level and rainfall readings at 15 minute intervals, plus 5000 sample events
Operating Temperature -18°C to 60°C (0°F to 140°F)
Operating Temperature -30°C to 60°C (-22°F to 140°F)
Power Two 6 volt lantern batteries (alkaline recommended) or one rechargeable 12 volt Isco 947 Lead Acid Battery
Program Memory Non-volatile, programmable flash; can be updated via Interrogator port without opening enclosure
Range (distance from sensor to liquid) Minimum 0.6m (2 ft); Maximum 3.7m (12 ft)
Sampler Activation Conditions Enabled, disabled, level, flow rate, rainfall and time; AND and OR combinations of any two of level, flow rate, rainfall and time; values may be above or below a set-point, inside or outside a range, or a rate of change
Sampler Pacing Output 12 volt pulse
Serial Connection to IBM PC® or compatible computer with Isco Flowlink Software
Size 26.7cm × 22.9cm × 15.2cm (10.5" × 9.0" × 6.0")
Span 0 to 3m (0 to 10 ft)
Storage Interval 1, 2, 5, 10, 15, 30, 60 or 120 minutes
Storage Modes Rollover, slate or triggered slate
Storage Temperature -40°C to 60°C (-40°F to 140°F)
Temperature Coefficient Maximum error over compensated temperature range (per degree of temperature change) ±0.000085 × D per °C (±0.000047 × D per °F) where D is the distance from the transducer to the liquid surface
Time Base Accuracy ±1 second per day
Typical Battery Life (1 hour data storage interval) With alkaline lantern batteries - 12 months
With Isco 947 Lead Acid Battery - 4 months (between recharges)
Battery life is proportionately less with more frequent data storage, eg, 3 months for 15 minute storage inteval with alkaline batteries
Weight (including cable) 1.2kg (2.6 lbs)
Weight (without batteries) 3.6kg (8.0 lbs)
Weirs V-notch, rectangular with and without end contractions, Cipolletti