AP-300 Automatic Polarimeter, Type A Temperature Controlled Package

Part Number: 5296Product Details

Overview

The AP-300 fully automatic polarimeter measures a sample easily by setting the observation tube (filled with a liquid sample) in the sample chamber and pressing START.

Digitally displayed measurement values are large and easy to read. The AP-300 generally used in pharmaceutical, perfume, cosmetic, industrial, chemical, and food factories is the best for measuring the angle of rotation. In addition, the built in International Sugar Scale is used widely in sugar factories.

ATAGO has prepared 2 different AP-300 packages for the sugar industry. Users have a choice between 2 types: Temperature Controlled Type A and Temperature Compensation Type B. The AP-300 comes with Automatic Temperature Compensation for International Sugar Scale. A Circulating Constant Temperature Bath is not necessary when the temperature compensation function is used or when the AP-300 is used in a temperature-controlled room.

• AP-300 (including accessories)
• Jacketed flow tube with funnel and temperature sensor for jacket (200mm - approximate volume 15ml)
* Circulating Constant Temperature Bath 60-C3 (Cat.No.1921) sold separately.
Features

• Automatic Temperature Compensation
• Can be connected to a printer
• RS-232C interface
• Digital display
• AC outlet (AC100 to 240V)

Specifications

Part Number: 5296
Dimensions 48.5cm × 28.5cm × 17.5cm
Display Panel Color LCD with backlight
Manual Calibration Enables manual calibration by measuring a standard Quartz Plate
Measurement Accuracy Angle of Rotation - 0.01°
International Sugar Scale
±0.03°Z (Checked by reading a standard Quartz Plate)
Measurement Range Angle of Rotation - -89.99 to +89.99°
International Sugar Scale - -130.00 to +130.00°Z
Measurement Readings Angle of Rotation, International Sugar Scale (without temperature compensation), International Sugar Scale (with Automatic Temperature Compensation ), Specific Optical Rotation, Concentration, and Purity
Measurement Scales Angle of Rotation , International Sugar Scale (without temperature compensation), International Sugar Scale (with Automatic Temperature Compensation)
Measurement Wavelength 589nm(D-line)
Minimum Indication Angle of Rotation - 0.01°
International Sugar Scale - 0.01°Z
Output Terminals Digital Printer DP-63 or DP-AD (optional)
RS-232C
Power Consumption 50VA
Power Supply AC100V to 240V, 50/60 Hz
Temperature correction range for International Sugar Scale 18.0 to 30.0°C
Weight 13.9kg