870B Baratron® Single-Ended Ultraclean Absolute Pressure Transducer and Local Display

Part Number: 870BProduct Details

Overview

Baratron® 800 Series transducers are ideally suited for use in delivery systems that feed ultrapure gases to critical process systems. Their wetted surfaces exposed to the gas stream have a finish better than 5 uin Ra. These transducers exhibit superior dry-down characteristics, and contribute no particles above background. After manufacture and assembly, they are purged with ultraclean nitrogen prior to double bagging.

The all-Inconel construction of the sensors in the Baratron® 800 Series transducers allows for high overpressure tolerances that reduce errors due to line pressure spikes. The 870 offers ±1% of Reading accuracy.

These transducers are suitable for use with SDS® (Safe Delivery Source) Systems.
Features

• Capacitance technology at a competitive price
• Industry's smallest integrated transducer with display <5"H
• Incoloy® wetted surface provides service superior corrosive gas and liquid compatability
• Rated for Class I, Division II service
• Highest overpressure ratio tolerances on the market assure no degradation in zero repeatability or performance
• Superior accuracy in lower pressure ranges vs competitive brands
• CE Mark compliant
• Optional NEMA 4 enclosure for outdoor
• 5 Ra finish (standard)
• Replaceable electronics

Specifications

Part Number: 870B
Accuracy (including non-linearity, hysteresis, and non-repeatability) 1.0% of Reading
Ambient Operating Temperature 0°C to 50°C (32° to 12°F)
Burst Pressure 10× F.S. for ranges from 10 Torr to 500 psi
5× F.S. for ranges from 501 to 3000 psi
Electrical Connectors Male 9-pin or male 15-pin Type "D" at end of 9" flying lead, Bendix® at end of 9" flying lead, or 6' or 10' flying leads
Electromagnetic Compatibility Fully CE Compliant to EMC Directive 89/336/EEC²
Fittings ½" weld stub, ¼" weld stub, ¼" butt-weld "T", 4 VCR male, 4 VCR female
Full Scale Ranges 1000 Torr to 3000 psia
Input Power Required 0 to 10 Volt output - +13 VDC to +32 VDC @ 10 mA max.
0 to 5 Volt output - +12 VDC to +32 VDC (regulated if below 13VDC) @ 10 mA max.
2-wire 4-20 mA - output +13 to +32 VDC excitation
Materials Exposed to Gases Incoloy®
Output 0 to 10 VDC - into > 10 Kohm load
0 to 5 VDC - into > 10 Kohm load
2-wire 4-20 mA output with +13 to +32 VDC at transducer terminals - Typically into 0 to 900W load (depending upon excitation)
Overpressure Limit¹ 1.5× F.S. for ranges from 1000 Torr to 500 psi
1.2× F.S. for ranges from 501 to 3000 psi
Temperature Coefficients - Span 0.04% of Rdg./°C, 50 psia to 3000 psia
0.08% of Rdg./°C, 1000 Torr
Temperature Coefficients - Zero 0.02% of F.S./°C, 50 psia to 3000 psia
0.04% of F.S./°C, 1000 Torr
Wetted Surfaces < 5µin Ra max.

Related Items

  • 223B Baratron® Differential Capacitance Manometer (±15 VDC Input Power)
  • 225A Baratron® Differential Capacitance Manometer (+24 VDC Input Power)
  • 890B Micro-Baratron® Single-Ended Absolute Ultraclean Pressure Transducers with Optional Local Display
  • 891B Micro-Baratron® Single-Ended Compound Ultraclean Pressure Transducers with Optional Local Display
  • 892B Micro-Baratron® Flow-Through Absolute Ultraclean Pressure Transducers with Optional Local Display
  • 893B Micro-Baratron® Flow-Through Compound Ultraclean Pressure Transducers with Optional Local Display
  • 872B Baratron® Flow-Through Ultraclean Absolute Pressure Transducer and Local Display
  • LDM Local Display Module for MKS 700 and 800 Series
  • LDM-A Local Display Module for MKS 700 and 800 Series, 4-20mA Current version
  • 41A Single-Ended Gauge Switch
  • 51A Single-Ended Absolute Vacuum/Pressure Switch
  • 42A Flow-Through Gauge Switch
  • 52A Flow-Through Absolute Vacuum/Pressure Switch
  • 141A Absolute Ambient Pressure/Vacuum Switch
  • 142A Absolute Pressure/Vacuum Switch (Temperature Regulated to 100°C)
  • 1651C Displayless Self-Tuning/Digital PID Pressure Controller
  • PDR2000 Digital Power Supply and Readout
  • 660B Single-Channel Digital Power Supply and Readout
  • PR4000 Digital Power Supply and Readout
  • 146C Vacuum Gauge Measurement and Control System
  • 154A High Flow Control Valve
  • 1249A Solenoid Control Valve Driver
  • 120AA Baratron® High Accuracy Absolute Capacitance Manometer (Integrated Signal Conditioner)
  • 120AD Baratron® High Accuracy Differential Capacitance Manometer (Integrated Signal Conditioner)
  • 148J All Metal Control Valve